Method and device for measuring electromagnetic shielding effectiveness of high-voltage shielded cable

文档序号:1183568 发布日期:2020-09-22 浏览:14次 中文

阅读说明:本技术 高压屏蔽电缆电磁屏蔽效能的测量方法和测量装置 (Method and device for measuring electromagnetic shielding effectiveness of high-voltage shielded cable ) 是由 李立嘉 沈涛 崔强 石磊 夏志立 于 2020-07-13 设计创作,主要内容包括:本发明公开了一种高压屏蔽电缆电磁屏蔽效能的测量方法,在30MHz-1GHz频段,按照如下步骤进行测量:S1:在电缆段连接一根已知插入损耗的同轴电缆以代替被测屏蔽电缆,在第一电流探头所在的被测频点测量功率值,在第二电流探头所在的被测频点测量功率值,消除系统测量误差;S2:在电缆段连接被测屏蔽电缆,测量第二电流探头所在的被测频点的功率值得到P<Sub>1</Sub>;S3:同时移动功率吸收钳和辅助吸收钳,找到被测频率点的最大值P<Sub>2</Sub>;S4:屏蔽电缆在该频点的屏效值为SE=P<Sub>1</Sub>-P<Sub>2</Sub>。在9KHz-30MHz频段采用另一种测量步骤。该方法能够准确测量高压屏蔽电缆的电磁屏蔽效能。(The invention discloses a method for measuring electromagnetic shielding effectiveness of a high-voltage shielded cable, which comprises the following steps of: s1: connecting a coaxial cable with known insertion loss to the cable section to replace a tested shielded cable, measuring a power value at a tested frequency point where the first current probe is positioned, measuring a power value at a tested frequency point where the second current probe is positioned, and eliminating a system measurement error; s2: connecting the tested shielded cable to the cable section, and measuring the power value of the tested frequency point where the second current probe is located to obtain P 1 (ii) a S3: simultaneously moving the power absorption clamp and the auxiliary absorption clamp to find the maximum value P of the measured frequency point 2 (ii) a S4: the effective shielding value of the shielded cable at the frequency point is SE ═ P 1 ‑P 2 . Another measurement procedure is adopted in the frequency band of 9KHz-30 MHz. The method can accurately measure the electromagnetic shielding effectiveness of the high-voltage shielded cable.)

1. The method for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable is characterized in that a measuring device is established in a frequency band of 30MHz-1GHz as follows:

arranging a signal generator and a terminal load at two ends, sequentially connecting a first current probe, a wall-penetrating base, a cable section and a second current probe from the signal generator to the terminal load, connecting a movable auxiliary absorption clamp and a power absorption clamp on the cable section, and connecting the power absorption clamp with a receiver; arranging a reflection grounding plate at the through-wall base;

the measurement was carried out as follows:

s1: connecting a coaxial cable with known insertion loss to the cable section to replace a tested shielded cable, measuring a power value at a tested frequency point where the first current probe is positioned, measuring a power value at a tested frequency point where the second current probe is positioned, and eliminating a system measurement error;

s2: connecting the screen to be tested to the cable sectionThe shielded cable measures the power value of the measured frequency point where the second current probe is positioned to obtain P1

S3: simultaneously moving the power absorption clamp and the auxiliary absorption clamp to find the maximum value P of the measured frequency point2

S4: the effective shielding value of the shielded cable at the frequency point is SE ═ P1-P2

2. The method for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable according to claim 1, wherein a current probe calibration device is further disposed at the first current probe and the second current probe.

3. The method for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable according to claim 1, wherein an electric measuring guide rail is further arranged beside the cable section.

4. The method for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable according to claim 1, wherein the measuring device is established in the frequency band of 9KHz to 30MHz as follows:

arranging a signal generator and a terminal load at two ends, and sequentially connecting a wall penetrating base, a third current probe, a cable section and a fourth current probe from the signal generator to the terminal load; arranging a reflection grounding plate at the through-wall base;

the measurement was carried out as follows:

s1: connecting a coaxial cable with known insertion loss to the cable section to replace a tested shielded cable, measuring a power value at a tested frequency point where the third current probe is positioned, measuring a power value at a tested frequency point where the fourth current probe is positioned, and eliminating a system measurement error;

s2: the cable section is connected with the tested shielded cable, the power value is measured at the tested frequency point where the third current probe is positioned, the power value is measured at the tested frequency point where the fourth current probe is positioned, and P is obtained1

S3: moving the position of the third current probe or the fourth current probe, measuring at three positions of the measured shielded cable, and averaging to obtain the average valueTo P2Or finding the maximum leak value as P2

S4: the effective shielding value of the shielded cable at the frequency point is SE ═ P1-P2

5. The method for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable according to claim 4, wherein a current probe calibration device is further disposed at the third current probe and the fourth current probe.

6. A device for measuring electromagnetic shielding effectiveness of a high-voltage shielded cable, which is suitable for the measuring method according to claim 1, and is characterized by comprising, in a frequency band of 30MHz-1 GHz: the device comprises a signal generator and a terminal load which are positioned at two ends, wherein a first current probe, a wall penetrating base, a cable section and a second current probe are sequentially connected from the signal generator to the terminal load; and a reflection grounding plate is arranged at the through-wall base.

7. The apparatus for measuring the electromagnetic shielding effectiveness of a high-voltage shielded cable according to claim 6, wherein the apparatus further comprises a signal generator and a terminal load disposed at two ends of the apparatus, wherein the signal generator and the terminal load are sequentially connected to the wall base, the third current probe, the cable segment and the fourth current probe from the signal generator to the terminal load; and a reflection grounding plate is arranged at the through-wall base.

Technical Field

The invention relates to a technology for testing electromagnetic shielding effectiveness of a high-voltage shielded cable, in particular to a method and a device for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable.

Background

The high-voltage cable is used for transmitting large currents such as direct-current voltage, power frequency voltage and the like. In the transmission switching process, a large number of radio frequency interference signals can be generated, severe electromagnetic pollution is caused to the use environment, in order to inhibit electromagnetic interference, a shielding layer is added outside a high-voltage cable, and the high-voltage shielding cable is formed and is widely applied to the fields of electric automobiles, power communication, aerospace, naval vessel military industry and the like. How to evaluate the shielding effectiveness of a high-voltage shielded cable is a problem of intense research, and the surface transfer impedance of the high-voltage shielded cable is measured by a line injection method or a triple-axis method to represent the shielding effectiveness of the cable. But encounter many problems and difficulties in implementation.

From the measurement port impedance classification, shielded cables can be classified into two types, fixed impedance shielded cables (e.g., 50 Ω coaxial cables, with impedance maintained at 50 Ω over a frequency range) and non-fixed impedance shielded cables (e.g., various shielded wires). Fixed impedance shielded cables are typically used for radio frequency signal transmission, while non-fixed impedance shielded cables are typically used for high voltage power transmission.

A method for measuring the electromagnetic shielding effectiveness of a fixed-impedance shielded cable, such as the power absorption clamp method, is to obtain a reference measured value P by using the characteristic that the impedance of the measured shielded cable is not changed (the whole measuring system is a 50 omega system)1. The maximum value P of the leakage radio frequency signal of the tested shielded cable is searched on a 6000mm test guide rail by using power absorption pliers2. Effective shielding value SE (equal to P) of shielded cable at measuring frequency point1-P2. The measurable frequency range is 30MHz-1 GHz.

The electromagnetic shielding effectiveness of an unfixed impedance shielded cable cannot be measured by using a method for measuring the electromagnetic shielding effectiveness of a fixed impedance shielded cable. The port impedance of the non-fixed impedance shielded cable can change along with frequency, length, thickness, manufacturing process and the like. Radio frequency signals are reflected at junctions where impedance discontinuities occur. How much the RF signal is reflected is not clear enough to obtain an accurate reference measurement P1Obtaining no P1A measurement of the electromagnetic shielding effectiveness cannot be achieved. This is why up to now there is no international oneA uniformly accepted measurement method is used to measure the cause of electromagnetic shielding effectiveness of non-fixed impedance shielded cables.

Disclosure of Invention

The invention aims to provide a method and a device for measuring the electromagnetic shielding effectiveness of a high-voltage shielded cable, which are used for accurately measuring the electromagnetic shielding effectiveness of a non-fixed impedance shielded cable.

In order to achieve the above object, in a first aspect, the present invention provides a method for measuring electromagnetic shielding effectiveness of a high-voltage shielded cable, wherein a measuring device is established in a frequency range of 30MHz to 1GHz as follows:

arranging a signal generator and a terminal load at two ends, sequentially connecting a first current probe, a wall-penetrating base, a cable section and a second current probe from the signal generator to the terminal load, connecting a movable auxiliary absorption clamp and a power absorption clamp on the cable section, and connecting the power absorption clamp with a receiver; arranging a reflection grounding plate at the through-wall base;

the measurement was carried out as follows:

s1: connecting a coaxial cable with known insertion loss to the cable section to replace a tested shielded cable, measuring a power value at a tested frequency point where the first current probe is positioned, measuring a power value at a tested frequency point where the second current probe is positioned, and eliminating a system measurement error;

s2: connecting the tested shielded cable to the cable section, and measuring the power value of the tested frequency point where the second current probe is located to obtain P1

S3: simultaneously moving the power absorption clamp and the auxiliary absorption clamp to find the maximum value P of the measured frequency point2

S4: the effective shielding value of the shielded cable at the frequency point is SE ═ P1-P2

Further, a current probe calibration device is arranged at the first current probe and the second current probe.

Furthermore, an electric measuring guide rail is arranged beside the cable section.

Further, in the frequency band of 9KHz-30MHz, a measuring device is established according to the following modes:

arranging a signal generator and a terminal load at two ends, and sequentially connecting a wall penetrating base, a third current probe, a cable section and a fourth current probe from the signal generator to the terminal load; arranging a reflection grounding plate at the through-wall base;

the measurement was carried out as follows:

s1: connecting a coaxial cable with known insertion loss to the cable section to replace a tested shielded cable, measuring a power value at a tested frequency point where the third current probe is positioned, measuring a power value at a tested frequency point where the fourth current probe is positioned, and eliminating a system measurement error;

s2: the cable section is connected with the tested shielded cable, the power value is measured at the tested frequency point where the third current probe is positioned, the power value is measured at the tested frequency point where the fourth current probe is positioned, and P is obtained1

S3: moving the position of the third current probe or the fourth current probe, measuring at three positions of the measured shielded cable, and averaging to obtain P2Or finding the maximum leak value as P2

S4: the effective shielding value of the shielded cable at the frequency point is SE ═ P1-P2

Further, a current probe calibration device is arranged at the third current probe and the fourth current probe.

In a second aspect, the present invention further provides a device for measuring electromagnetic shielding effectiveness of a high-voltage shielded cable, which is suitable for the above measuring method, and includes: the device comprises a signal generator and a terminal load which are positioned at two ends, wherein a first current probe, a wall penetrating base, a cable section and a second current probe are sequentially connected from the signal generator to the terminal load; and a reflection grounding plate is arranged at the through-wall base.

Furthermore, the frequency range of 9KHz-30MHz, the measuring device also comprises a signal generator and a terminal load which are arranged at two ends, and the signal generator and the terminal load are sequentially connected with the wall-through base, the third current probe, the cable section and the fourth current probe; and a reflection grounding plate is arranged at the through-wall base.

Compared with the prior art, the method can accurately measure the reference value P of the non-fixed impedance shielded cable by a reasonably-built measuring system and applying an innovative measuring method1And finally obtaining the electromagnetic shielding effectiveness of the non-fixed impedance shielded cable on the basis.

Drawings

FIG. 1 is a diagram of a vector network analyzer test connection;

FIG. 2 is a diagram showing the test results of the impedance value of a 0.6m thin wire cable;

FIG. 3 is a diagram showing the test results of attenuation values of 0.6m thin wire cables;

FIG. 4 is a diagram showing the results of the test of the impedance value of a 1.2m thin wire cable;

FIG. 5 is a graph showing the test results of attenuation values of 1.2m thin wire cables;

FIG. 6 is a diagram showing the results of the test of the impedance value of a thin wire cable of 6 m;

FIG. 7 is a graph showing the test results of attenuation values of a 6m thin wire cable;

FIG. 8 is a diagram of the results of testing the impedance of a 0.6m thick wire cable;

FIG. 9 is a graph showing the test results of attenuation values of 0.6m thick cable;

FIG. 10 is a graph showing the results of testing the impedance of a 1.2m thick wire cable;

FIG. 11 is a graph showing the test results of attenuation values of 1.2m thick-line cables;

FIG. 12 is a diagram showing the results of testing the impedance of a 6m thick-wire cable;

FIG. 13 is a graph showing the test results of attenuation values of cables with 6m thick wires;

fig. 14 is a device (30MHz-1GHz band) for measuring electromagnetic shielding effectiveness of a high-voltage shielded cable according to an embodiment of the present invention;

FIG. 15 is a schematic diagram of an apparatus for measuring electromagnetic shielding effectiveness of a high-voltage shielded cable (frequency band of 9kHz-30MHz) according to an embodiment of the present invention;

FIG. 16 is a graph of effective value test curve (9kHz-30MHz) for a 6m thin wire cable screen;

FIG. 17 is a graph showing the screen effective value test curve (30MHz-300MHz) of a 6m thin wire cable.

Detailed Description

The principles and spirit of the present invention will be described with reference to a number of exemplary embodiments shown in the drawings. It should be understood that these embodiments are described only to enable those skilled in the art to better understand and to implement the present invention, and are not intended to limit the scope of the present invention in any way.

Before describing the technical solutions of the embodiments of the present invention, the present invention is first described in the research and analysis, which are important for the present invention, and the inventors have made creative efforts for this purpose.

Reference measurement P needed by considering how to obtain non-fixed impedance shielded cable screen effect measurement1In the invention, the characteristics and the change rule of the vector network analyzer are firstly researched by the vector network analyzer. Fig. 1 is a test connection diagram. Fig. 2 to 7 show the results of impedance and attenuation tests for different lengths of the same shielded cable. Fig. 8 to 13 show the results of impedance and attenuation tests for different lengths of another shielded cable. The relevant test data are shown in tables 1-6.

TABLE 10.6 m Fine line impedance and attenuation values

Frequency of Impedance (L) Attenuation value
30kHz 49Ω 0dB
50kHz 49Ω 0dB
494kHz 50Ω 0dB
1MHz 50Ω -1dB
10MHz 44Ω 0dB
57MHz 14Ω -2dB
140MHz 49Ω 0dB
226MHz 23Ω -3dB
286MHz 49Ω -2dB
363MHz 30Ω -3dB
400MHz 49Ω -2dB

TABLE 21.2 m Fine line impedance and attenuation values

Frequency of Impedance (L) Attenuation value
30kHz 49Ω 0dB
1MHz 49Ω 0dB
10MHz 33Ω 0dB
37MHz 13Ω -3dB
69MHz 44Ω -1dB
110MHz 17Ω -3dB
147MHz 37Ω -1dB
187MHz 24Ω -3dB
215MHz 44Ω -2dB
248MHz 21Ω -4dB
286MHz 46Ω -2dB
330MHz 30Ω -4dB
363MHz 44Ω -3dB
400MHz 31Ω -4dB

TABLE 36 m Fine line impedance and attenuation values

Figure BDA0002581720020000061

TABLE 40.6 m bold line impedance and attenuation values

Frequency of Impedance (L) Attenuation value
30kHz 50Ω 0dB
1Mhz 50Ω 0dB
9Mhz 24Ω -2dB
39Mhz -7dB
75Mhz 44Ω -1dB
116Mhz -7dB
154Mhz 36Ω -1dB
196Mhz 11Ω -6dB
226Mhz 42Ω -2dB
273Mhz 13Ω -7dB
300Mhz 42Ω -2dB
346Mhz 14Ω -7dB
381Mhz 37Ω -2dB
400Mhz -5dB

TABLE 66 m Thick line impedance and attenuation values

Frequency of Impedance (L) Attenuation value
30kHz 49Ω 0dB
518kHz 45Ω 0dB
1MHz 35Ω -1dB
7MHz -6dB
14MHz 38Ω -1dB
22MHz -7dB
27MHz 36Ω -2dB
35MHz -7dB
42MHz 32Ω -2dB
49MHz -7dB
57MHz 30Ω -2dB
62MHz -7dB
72MHz 26Ω -3dB
49MHz -7dB
87MHz 21Ω -4dB
237MHz 10Ω -9dB
273MHz 21Ω -7dB
381MHz 14Ω -10dB

As can be seen from fig. 2-13: different cables or the same cable length will have the following characteristics:

the impedance of the shielded cable varies and is large with the change of frequency.

② the same shielded cable length is different and the impedance is also different and the variation is large.

And thirdly, when the geometric dimension of the same shielded cable is fixed, the impedance changes only as a function of frequency (single variable).

Fourthly, when the geometric dimension, the structural form, the measuring frequency and other factors of the cable are determined, the impedance is a fixed value, and the attenuation is a fixed value.

The same shielded cable has no linear relation between length and attenuation (at some frequency points, the insertion loss is not always large when the long line is long)

Sixthly, the impedance varies greatly with the frequency, but is less than 50 omega.

The impedance of the cable below 1MHz is basically unchanged.

That is, when a shielded cable of a certain length and a non-fixed impedance is placed in a 50 Ω measuring system, the desired reference measured value P can be obtained accurately by measuring the attenuation1The measured shielded cable can be regarded as a radio frequency attenuator changing along with the change of frequency in a 50 omega measuring system, and the change of the S21 value really reflects a measurement reference value P caused by the change of frequency, impedance and reflection of a radio frequency signal in the measured shielded cable1The difference between the change of the reference value P and the maximum screen effect measurement method of the fixed impedance shielded cable is that the reference value P is the reference value P in the screen effect measurement of the non-fixed impedance shielded cable1The measurement is changed according to the type, length, frequency and the like of the cable, and the measurement is needed from time to time (the reference value is greatly changed, and double measurement errors are brought).

Accordingly, the embodiment of the invention puts the shielded cable into a 50 omega measuring system, and the reference value P of the screen effect measurement can be obtained by measuring the attenuation1Two sets of devices are designed for dividing the measuring frequency of 9kHz-1GHz into two parts of 30MHz-1GHz and 9kHz-30MHz respectively to measure the screen effect, so that the attenuation value P can be obtained2

The 30MHz-1GHz cable screen effect measuring device and method:

the frequency band of 30MHz-1GHz is measured by the measuring device of FIG. 14. Two current probe calibration devices are connected in series on the basis of the original device for measuring the screen effect of the 50 omega coaxial cable, and on the basis of ensuring that the whole measuring system (except the measured shielded cable) is a 50 omega system in the measuring frequency band, the current probe is utilized to obtain the reference value P of the measuring signal1The attenuation measurement P is obtained by finding the maximum value of the leakage signal on a 600mm measuring rail by means of power absorption clamps and coupling clamps2So as to obtain the screen effect measured value SE-P of the shielded cable1-P2

Specifically, a signal generator 11 and a terminal load 17 are arranged at two ends, a first current probe 13a, an N-N wall socket 18, a cable section and a second current probe 13b are sequentially connected from the signal generator 11 to the terminal load 17, a movable auxiliary absorption clamp 14 and a power absorption clamp 15 are connected to the cable section, and the power absorption clamp 15 is connected to a receiver 110. A 2m multiplied by 2mm grounded reflection copper plate 19 is arranged at the through-wall base 18, an electric measuring guide rail 16 with the length of 6m is arranged beside the cable section, and a current probe calibration device 12 is arranged at the first current probe 13a and the second current probe 13 b. Wherein, the cable section is used for arranging the tested shielded cable 111 or a cable for eliminating system measurement errors.

The measurement method is carried out as follows:

s1: a50 omega coaxial cable with known insertion loss is used for replacing the tested shielded cable 111, and power values of two points are measured A, D at the tested frequency point, so that system measurement errors are eliminated.

S2: connecting the tested shielded cable 111, and measuring the power value of the D point on the tested frequency point to obtain P1

S3: simultaneously moving the power absorption clamp and the auxiliary absorption clamp to find the maximum value P of the measured frequency point2

S4: the effective shielding value of the shielded cable at the frequency point is SE ═ P1-P2

The shielding value of the shielded cable obtained by the method is independent of the length of the shielded cable in the frequency band of 30MHz-1GHz, and the method is recommended to be carried out in a shielded room.

The 9kHz-30MHz cable screen effect testing device and method comprises the following steps:

and measuring the shielding effectiveness of the 9KHz-30MHz frequency band shielding cable. In principle, similar methods can be applied, but the practical implementation is difficult. The wavelength of 30MHz is 10 meters, the reference value and the maximum leakage value of the measuring frequency of the shielded cable above 30MHz are found in the half wavelength, but the wavelength of 10MHz is 30 meters, the wavelength of 3MHz is 100 meters, the wavelength of 1MHz is 300 meters … …, the wavelength of 10KHz is 30000 meters, and then P is found in the half wavelength1And P2It is difficult.

As discussed above, the characteristic impedance of the shielded cable in the frequency band below 1MHz has a small change, substantially 50 Ω (in the case of a termination load of 50 Ω), that is, the measurement uncertainty of the shielded cable below 1MHz caused by reflection is small and can be ignored, if the maximum leakage of the shielded cable with half wavelength is not to be found, half of the wavelength of 1MHz is taken, and the maximum leakage is not to be found in the frequency band below 1 MHz.

Because the length of the same type of shielded cable is different, the impedance is different, and the reflection is different. The effective value of the measurement screen will also be different. Therefore, the screen effective value of the shielded cable is measured in the frequency range of 9KHz-30MHz, and the measurement is carried out based on the length of the shielded cable actually used. If the manufacturer of the shielded cable has conditions (can build a measuring environment with half wavelength), the screen effective value of the shielded cable can be measured by the same measuring method.

It should be noted that, it is most accurate to measure the screen effective value of the half-wavelength shielded cable by using this measurement method, (for this shielded cable line as a whole), and it is most suitable for the actual use (for the user of the shielded cable) to measure the screen effective value of the shielded cable with the actual used length.

The screen effect measuring connection of the 9KHz-30MHz frequency range shielded cable is as shown in FIG. 15, a signal generator 21 and a terminal load 26 are arranged at two ends, and an N-N wall penetrating base 22, a third current probe 25a, a cable section and a second current probe 25b are sequentially connected from the signal generator 21 to the terminal load 26; a 2m multiplied by 2m grounding reflection copper plate 23 is arranged at the through-wall base; the third current probe 25a and the fourth current probe 25b are also provided with a current probe calibration device 24, and are also connected with a grounding copper plate 28. The device is also equipped with a 0.8m high non-metallic test stand 27.

The measurement was carried out as follows:

s1: a50 omega coaxial cable with known insertion loss is used for replacing a tested shielded cable, and power values of two points are measured A, E at a tested frequency point, so that system measurement errors are eliminated.

S2: connecting the tested shielded cable, and measuring the power values of the two points E on the tested frequency point to obtain P1

S3: the position of the current probe 25 (the current probe 25 may be the third current probe 25a or the fourth current probe 25b) is moved, the three positions are B, C, D measured, and P is obtained by averaging2(or finding the maximum leakage value P2)

S4: the effective shielding value SE of the shielded cable at the frequency point is equal to P1-P2

Test results

Tables 7 and 8 are screen effective values measured when the same shielded cable is different in length;

the table 9 and the table 10 are screen effective values measured when the other shielding cables of the same type have different lengths;

table 11 shows the screen effective values measured for another shielded cable.

And (4) setting the screen effective value SE as the E point value- (the maximum value of the B point value, the C point value and the D point value).

TABLE 70.6 thick line test results

Date: wire length: 0.6m type: output amplitude of thick line source of 110dBuV

Frequency of A point value E point value B point value C point value D point value Screen effective value
10kHz
110 108 81 81 82 26
50kHz 112 104 71 71 71 33
100kHz 110 106 67 67 67 39
500kHz 107 105 55 55 55 50
1MHz 105.7 103.6 50.6 50.6 50.6 53
5MHz 106.4 104.2 57.4 56.4 56.4 46.8
10MHz 109 103.2 52.3 52.3 52.3 50.9
20MHz 111 100.7 52.8 52.8 52.8 47.9
30MHz 111.4 99 51.3 50.3 51.3 47.7

TABLE 81.2 m thick line test results

Date: wire length: 1.2m type: output amplitude of thick line source of 110dBuV

Frequency of A point value E point value B point value C point value D point value Screen effective value
10kHz 107 107 79 79 79 28
50kHz 109 106 65 66 66 40
100kHz 108 102 62 62 62 40
500kHz 104 100 51 51 51 49
1MHz 100.7 97.6 45.6 40.6 46.6 51
5MHz 95.2 90.4 54.4 51.4 54.4 40.8
10MHz 106.2 91 46.3 46.3 46.3 59.9
20MHz 95.7 92 48.8 53.8 47.8 41.9
30MHz 107.2 96.7 54.3 54.3 54.3 42.4

TABLE 90.6 m Fine line test results

Date: wire length: 0.6m type: output amplitude of thin line source of 110dBuV

Frequency of A point value E point value B point value C point value D point value Screen effective value
10kHz 108 108 86 86 86 22
50kHz 110 106 75 75 75 31
100kHz 110 106 72 72 72 34
500kHz 107 105 61 61 61 44
1MHz 105.7 103.6 56.6 56.6 56.6 47
5MHz 106.4 104.2 60.4 60.4 60.4 43.8
10MHz 108 104.2 57.3 58.3 58.3 45.9
20MHz 110 102.7 55.8 57.8 56.8 44.9
30MHz 110.4 101 54.3 53.3 54.3 46.7

TABLE 101.2 m Fine line test results

Date: wire length: 1.2m type: output amplitude of thin line source of 110dBuV

Frequency of A point value E point value B point value C point value D point value Screen effective value
10kHz 109 107 90 90 90 19
50kHz 110 106 78 78 78 28
100kHz 110 106 74 74 74 32
500kHz 107 104 61 61 61 43
1MHz 105.7 103.6 55.6 55.6 55.6 48
5MHz 107.4 104.2 57.4 57.4 57.4 46.8
10MHz 109 103.2 55.3 55.3 55.3 47.9
20MHz 111 100.7 55.8 54.8 55.8 44.9
30MHz 110.4 100 53.3 51.3 54.3 45.7

TABLE 116 m Fine line test results

Date: 12/2 line length: the model of 6m is: output amplitude of thin line source of 110dBuV

Frequency of A point value E point value C point value Screen effective value
30MHz 109.3 101 61 40
50MHz 101.7 100.6 68.1 32.5
100MHz 107.1 98.3 66.4 31.9
120MHz 108.2 97.5 65.1 32.4
150MHz 106 98.2 77.8 20.4
200MHz 105.3 90.3 71.1 19.2
250MHz 105 90 71.5 18.5
300MHz 100 87 64.5 22.5
400MHz 99 86 75.9 10.1

FIG. 16 is a screen virtual value test curve (9kHz-30MHz) of a 6m fine line cable, and FIG. 17 is a screen virtual value test curve (30MHz-300MHz) of the 6m fine line cable.

It must be noted that: the joint part of the tested shielded cable must adopt strict shielding measures. And ensuring that the screen effective value of the part is higher than that of the tested shielded cable.

The method for measuring the electromagnetic shielding effectiveness of the high-voltage shielded cable, namely the current probe method, provided by the embodiment of the invention is proved to be accurate in test data and good in repeatability through tests and verifications for several months. The problem of measuring the electromagnetic shielding effectiveness of the high-voltage shielding cable in the frequency range of 9kHz-400MHz is basically solved, and the measuring frequency range can be expanded to 9kHz-1 GHz. The method is also suitable for testing the shielding effectiveness of the coaxial cable.

The embodiment provides a novel method for measuring the electromagnetic shielding effectiveness of a high-voltage shielded cable, namely a current probe method. The method is based on the basic principle of electromagnetic shielding effectiveness measurement. The high-voltage shielded cable to be measured is put into a specially-arranged 50 omega measuring system. Through the current probe and the calibration device thereof, a reference measurement value P required by the measurement of the electromagnetic shielding effectiveness of the high-voltage shielded cable is accurately obtained in the frequency range of 9kHz-1GHz1. Obtaining a shielded attenuation measured value P by using a current probe at a frequency range of 9kHz-30MHz2Obtaining the shielded attenuation measured value P by using a power absorption clamp in the frequency range of 30MHz-1GHz2So as to measure the effective shielding value SE (SE is P) of the high-voltage shielded cable1-P2)

The method provides effective shielding indexes of the high-voltage shielding cable for further application of the high-voltage shielding cable in the fields of power communication, electric automobiles, aerospace, ship manufacturing and the like, and provides beneficial help for solving the problem of system-level electromagnetic compatibility.

The inventive concept is explained in detail herein using specific examples, which are given only to aid in understanding the core concepts of the invention. It should be understood that any obvious modifications, equivalents and other improvements made by those skilled in the art without departing from the spirit of the present invention are included in the scope of the present invention.

22页详细技术资料下载
上一篇:一种医用注射器针头装配设备
下一篇:环形缩距天线测试装置

网友询问留言

已有0条留言

还没有人留言评论。精彩留言会获得点赞!

精彩留言,会给你点赞!